Sciweavers

BILDMED
2009

3D Segmentation and Quantification of the Aortic Arch for Endovascular Aortic Repair

14 years 16 days ago
3D Segmentation and Quantification of the Aortic Arch for Endovascular Aortic Repair
Abstract. We introduce a novel model-based approach for the segmentation and quantification of the aortic arch morphology in 3D CTA images for endovascular aortic repair. The approach is based on a 3D analytic intensity model for thick vessels, which is directly fitted to the image. Based on the fitting results we compute the (local) 3D vessel curvature and torsion as well as the relevant lengths not only along the 3D centerline but particularly along the inner and outer contour. These measurements are important for pre-operative planning. We have successfully applied our approach using 3D CTA images and have compared the results with ground truth obtained by a radiologist. It turned out that our approach yields accurate estimation results.
Stefan Wörz, Hendrik von Tengg-Kobligk, Veren
Added 08 Nov 2010
Updated 08 Nov 2010
Type Conference
Year 2009
Where BILDMED
Authors Stefan Wörz, Hendrik von Tengg-Kobligk, Verena Henninger, Dittmar Böckler, Hans-Ulrich Kauczor, Karl Rohr
Comments (0)