Sciweavers

ISBI
2006
IEEE

Exact correction of sharply varying off-resonance effects in spiral MRI

15 years 8 days ago
Exact correction of sharply varying off-resonance effects in spiral MRI
Magnetic Resonance Imaging with non-Cartesian acquisition schemes suffer from blurring artifacts induced by offresonance. Conventional algorithms for off-resonance correction are approximate in nature and assume that the offresonance map is varying smoothly across the object. This assumption is violated in the case of susceptibility related off-resonance which is sharply varying near certain tissue interfaces. Iterative methods have been proposed for sharply varying off-resonance correction but are either computationally intense in the exact form or need approximations like time segmentation to reduce computational cost. Here a new method for accurate and computationally efficient off resonance correction using spatial domain deconvolution is presented.
Mahender K. Makhijani, Krishna S. Nayak
Added 20 Nov 2009
Updated 20 Nov 2009
Type Conference
Year 2006
Where ISBI
Authors Mahender K. Makhijani, Krishna S. Nayak
Comments (0)