Sciweavers

TMI
2008

Robust Linearized Image Reconstruction for Multifrequency EIT of the Breast

13 years 11 months ago
Robust Linearized Image Reconstruction for Multifrequency EIT of the Breast
Abstract--Electrical impedance tomography (EIT) is a developing imaging modality that is beginning to show promise for detecting and characterizing tumors in the breast. At Rensselaer Polytechnic Institute, we have developed a combined EIT-tomosynthesis system that allows for the coregistered and simultaneous analysis of the breast using EIT and X-ray imaging. A significant challenge in EIT is the design of computationally efficient image reconstruction algorithms which are robust to various forms of model mismatch. Specifically, we have implemented a scaling procedure that is robust to the presence of a thin highly-resistive layer of skin at the boundary of the breast and we have developed an algorithm to detect and exclude from the image reconstruction electrodes that are in poor contact with the breast. In our initial clinical studies, it has been difficult to ensure that all electrodes make adequate contact with the breast, and thus procedures for the use of data sets containing po...
Gregory Boverman, Tzu-Jen Kao, Rujuta Kulkarni, Bo
Added 15 Dec 2010
Updated 15 Dec 2010
Type Journal
Year 2008
Where TMI
Authors Gregory Boverman, Tzu-Jen Kao, Rujuta Kulkarni, Bong Seok Kim, David Isaacson, Gary J. Saulnier, Jonathan C. Newell
Comments (0)