Sciweavers

ICRA
2008
IEEE

An easy calibration for oblique-viewing endoscopes

14 years 7 months ago
An easy calibration for oblique-viewing endoscopes
— Oblique-viewing endoscopes (oblique scopes) are widely used in minimally invasive surgery. The viewing direction of an oblique endoscope can be changed by rotating the scope cylinder, which enables a larger field of view, but makes the scope calibration process more difficult. The calibration is a critical step for incorporating oblique scope into computer assisted surgical procedures (robotics, navigation, augmented reality), though few calibration methods of oblique endoscopes has been developed. Yamaguchi et al. [1] first modelled and calibrated the oblique scope. They directly tracked the camera head and formulated the scope cylinder’s rotation to the camera model as an extrinsic parameter. Their method requires five additional parameters to be estimated. In this work, we track the scope cylinder instead. Since the rotation of the camera head with respect to the cylinder only causes the rotation of the image plane, less parameter needs to be estimated. Experiments demonst...
Chenyu Wu, Branislav Jaramaz
Added 30 May 2010
Updated 30 May 2010
Type Conference
Year 2008
Where ICRA
Authors Chenyu Wu, Branislav Jaramaz
Comments (0)